Storage and Handling
WARNING! Environmental Hazard. Do not dispose of system components in municipal
waste. Follow local regulations when disposing of components.
The environmental requirements for the storage and transport of the ExionLC
TM
100 system:
• Altitude not exceeding 1 828 m (6 000 feet) above sea level
• An ambient temperature of 4 °C to 35 °C (39.2 °F to 95 °F)
The rate of temperature change must not exceed 2 °C (3.6 °F) per hour. Ambient temperature fluctuations
exceeding these limits will result in higher signal drift and an inconsistent baseline signal from the detector.
For ambient temperatures between 4 °C and 30 °C, a non-condensing relative humidity of between 20% and
85% is required.
For ambient temperatures between 30 °C and 35 °C, a non-condensing relative humidity of between 20% and
70% is required.
Note:
Do not place the system adjacent to heaters or cooling ducts, or in direct sunlight.