STATIM
Press Stop To Reset This message is displayed on all error faults. Press STOP to clear message.
Touch pads do not work
Disconnect keypad plug from PCB. Be sure blue plastic piece for keypad plug on PCB is pushed
up on the plug pins as far as possible. Reconnect keypad and check. Replace keypad if necessary.
No
display or garbled display on LCD.
Check plug connections from cover to PCB. Check that microprocessor is seated firmly in socket.
Statim makes a clicking noise with cassette removed
This is caused by steam leaking from the cassette. The steam gets into the micro switch causing
the contacts to open and close and the solenoid clicks. Repair cassette leak and clicking should
stop in approximately 24 hours.
Loud buzzing noise
Clean or replace solenoid as needed.
Noise during drying cycle only
Some check valve noise is normal. Check the air filter. Replace if dirty. If filter is wet replace
check valve and compressor if necessary.
Water dripping from drain tube under Statim
Replace seal or repair cassette as needed.
Steam is escaping from Condenser Bottle vent hole
Ensure that condenser bottle is always filled to MIN line.
Steam is leaking from Push-In-Fitting at rear of Statim
Ensure that exhaust tube is fully inserted in fitting. Push past initial resistance until
tube seats.
Wraps remain wet after drying