Cycle Fault 6
The software has detected a steam generator (boiler) temperature 5°C greater than the chamber,
within 7.2seconds after a purge during the sterilizing phase of a cycle.
Calibrate boiler. Refer to
Boiler Calibration Section
.
Cycle Fault 7
The cassette temperature has dropped 4°C below set point. If visible steam leaks appear replace
cassette seal, lid or tray as needed. Disassemble solenoid valve and check for debris. Make sure
plunger slides smoothly in plunger tube. Refer to Solenoid Valve Section.
Cycle Fault 8
The software has detected a steam generator (boiler) temperature 5'less than the chamber, within
7.2 seconds after a purge during the sterilizing phase ofa cycle. Calibrate boiler. Refer to Boiler
Calibration Section. Check for kinks in the exhaust tubing. Check to see if condenser bottle is full
of hot water.
Cycle Fault 9 N/A
Cycle Fault 10
The cassette temperature has failed to drop to 115°C during the Unwrapped or Wrapped Cycle or
the temperature has failed to drop to 110°C during the Rubber and Plastics Cycle in the purge
conditioning stage. Check for kinks in the exhaust tubing. Check for a clogged duct in the rear of
the cassette tray. Blow out with compressed air. Check solenoid valve for proper operation.
Cycle Fault 11
The cassette temperature has failed to drop to 102°C within 60 seconds of the end of a cycle
during venting. Check for kinks in the exhaust tubing. Check for a clogged duct in the rear of the
cassette tray. Blow out with compressed air. Check the solenoid valve for proper operation.
Cycle Fault 12