SciCan STATIM Скачать руководство пользователя страница 11

STATIM

the front left corner of the Statim 5000 Upward, and reinsert the tube into the clip. 
Push the excess length of tubing back from where it came

 

Changing the Cassette Seal 

 

The cassette seal will last approximately 500 to 800 cycles. There should never be steam leaking 
from the front of the Statim or water dripping from under the Statim. To change the seal follow 
these directions: 

1. Place the cassette lid and the new seal on a clean work surface.

 

2. Examine the position of the old seal in the cassette lid and arrange the new seal in the 
same orientation. 

 

3. Remove the old seal with a small screwdriver, sliding it behind the corner of the seal and 
pulling the seal from the channel. Clean any residue from the seal channel. 

 

4. Excessively lubricate the new seal using liquid hand soap (provided with new seal). 

 

5. Align the holes in the new seal with the holes in the lid.

 

6. Install the holes first, then install the four corners allowing the sides to hang out. Be sure 
that two indicator squares are visible in each corner. 

 

7. Starting in the middle of along side insert the rounded edge of the seal into the channel. 
Work from the middle to one corner, then from the middle to the opposite corner. Install all 
four sides the same way. 

 

8. Check to see that the seal is fully inserted into the channel on all four sides. Check the 
corners to see that two indicator squares are visible in each corner. Check the two holes to 
be sure that they are centered in the cutouts. 

 

9. If there is a small steam leak or water under the unit after installing the new seal try 
another cycle. It may take time to seat the new seal. If the leak persists, remove the seal and 
reinstall. If unit continues to leak after installing seal for the second time try a new cassette 
lid or tray. 

 

http://www.sterilizers.com/statim-trouble-shooting-manual.html (11 of 26) [3/27/2004 9:23:44 PM]

Содержание STATIM

Страница 1: ...f these conditions occur check your cassette and seal All leaks should be corrected before proceeding with troubleshooting procedures Check Cassette 1 Chamber temperature does not rise more than 2 C i...

Страница 2: ......

Страница 3: ......

Страница 4: ......

Страница 5: ...Calibration Section Check for kinks in the exhaust tubing Check to see if condenser bottle is full of hot water Cycle Fault 9 N A Cycle Fault 10 The cassette temperature has failed to drop to 115 C d...

Страница 6: ......

Страница 7: ......

Страница 8: ...open and close and the solenoid clicks Repair cassette leak and clicking should stop in approximately 24 hours Loud buzzing noise Clean or replace solenoid as needed Noise during drying cycle only So...

Страница 9: ...of cover until it clears internal components then slide cover forward about an inch Gently rotate the cover up and pivot it around the front left corner of the machine When the cover is clear of the m...

Страница 10: ......

Страница 11: ...holes in the new seal with the holes in the lid 6 Install the holes first then install the four corners allowing the sides to hang out Be sure that two indicator squares are visible in each corner 7 S...

Страница 12: ...imately 800 cycles and if maintained properly could last longer Pump Tester Instructions I Disconnect the Teflon boiler inlet tube from the top of the boiler 2 Connect the disconnected end of the inle...

Страница 13: ...he compression nut holding the steam generator boiler outlet tube to the steam generator boiler outlet fitting 8 Remove the two screws from the steam generator boiler bracket The steam generator boile...

Страница 14: ...ting Tighten finger tight 5 Connect the compression nut holding the steam generator boiler outlet tube to the probe bracket inlet fitting Tighten finger tight 6 Connect the compression nut holding the...

Страница 15: ...e bracket Trim away the excess RTV silicone and remove the screws Do Not remove the probe bracket to get at the screws If you have trouble getting at the screws disconnect the gray hose between the ch...

Страница 16: ...the reservoir is filled with good distilled water and the condenser bottle is not full 6 After 15 minutes have passed start an unwrapped cycle while simultaneously activating the pump You may operate...

Страница 17: ...d chamber temperatures The value in the lower right corner of the display is the chamber temperature F9 CALIBRATING C8 2 00 C 135 00 C EXAMPLE 5 To start the steam generator boiler self calibration cy...

Страница 18: ...f the Statim on Press the WRAPPED button again to scroll to the next language Press the UNWRAPPED button to scroll to the previous language When the desired language is displayed press the STOP button...

Страница 19: ...button Press the button repeatedly to select subsequent fields 7 To save changes and return to regular operating mode press the STOP button Diagnosing Water Quality Sensor Problems The water quality...

Страница 20: ......

Страница 21: ......

Страница 22: ......

Страница 23: ......

Страница 24: ......

Страница 25: ...x 7 1 2 Cassette Internal Dimensions 15 x 7 x 1 Reservoir Capacity 4 liters distilled water Approximately 20 cycles Weight Without Water 72 lbs Power Consumption 1300 W Optional Printer Type Thermal...

Страница 26: ...ber and Plastics 40 minutes Heavy Duty Unwrapped 12 minutes Warranty I Year Parts excluding gasket 1 Year SciCan Labor http www sterilizers com statim trouble shooting manual html 26 of 26 3 27 2004 9...

Отзывы: