10. Route the black wire from the fuse assembly to the PCB connector J1-3 and connect the wire.
11. Bundle the loose wires together, using nylon cable ties every 2-3 inches.
12. Carefully push the compressor tube onto the check valve inlet and secure
the tube to the valve using a high temperature application cable tie.
Microswitch Removal and Replacement
'To remove the microswitch follow these steps:
1. Turn the main power switch OFF, and unplug the unit.
2. Disconnect the microswitch leads from the PCB block terminal positions labeled
CASSIN and carefully cut the cable ties securing the microswitch leads.
3.There may be RTV silicone sealant on the heads of the two screws securing the
microswitch to the side of the probe bracket. Trim away the excess RTV silicone and
remove the screws. Do Not remove the probe bracket to get at the screws. If you have
trouble getting at the screws disconnect the gray hose between the check valve and the
compressor and remove the four compressor mounting screws. Set the compressor off the
back of the Statim being careful not to damage the wiring.
4. Remove the microswitch.
5. Remove all RTV silicone residue from the probe bracket surface.
To replace the microswitch follow these steps:
1. Attach the replacement microswitch to the probe bracket.
2.
Connect the microswitch leads to the PCB block terminal positions labeled CASSIN.
Secure the stripped leads.
3. Route the microswitch leads the same as before and secure with nylon cable ties.
4.
Plug in the power cord and turn on the main power switch. When the cassette is fully