STATIM

The Display Reads 

 

14:23 03/ 1 1 / 1998 

 

hh: mm dd/mm/yy 

 

EXAMPLE NOTE: The date reads day, month and year.

 

3. There are five fields on the display showing the hours (24 hour clock), minutes, day of 
the month and year. The cycle buttons of the keypad are used to select a field for setting and 
changing that field's value. The blinking cursor highlights the field currently selected. 

 

4. To increase the value of the selected field, press the UNWRAPPED cycle button. Hold 
the button down to continuously increase the value. 

 

5. To decrease the value of the selected field, press the WRAPPED cycle button. Hold the 
button down to continuously increase the value. 

 

6. To select the next field for setting, press the RUBBER AND PLASTICS cycle button. 
Press the button repeatedly to select subsequent fields.

 

7. To save changes and return to regular operating mode, press the STOP button. 

 

Diagnosing Water Quality Sensor Problems 

 

The water quality sensor detects whether or not there is sufficient water in the 
reservoir, monitors conductivity of the water in the reservoir and prevents a 
cycle from starting if the water measured is unsuitable. The allowed value of 
dissolved solids is 5 ppm or less. 

The water quality sensor can fail in one of four ways: 

1. The sensor always indicates an empty reservoir. 
2. The sensor always indicates a reservoir full of good quality water. 
3. The sensor always indicates a reservoir full of poor quality water. 
4. The sensor indicates random water quality readings. 

A water quality sensor failure is caused by: poor quality water (drain reservoir 

http://www.sterilizers.com/statim-trouble-shooting-manual.html (19 of 26) [3/27/2004 9:23:44 PM]

Содержание STATIM

Страница 1: ...f these conditions occur check your cassette and seal All leaks should be corrected before proceeding with troubleshooting procedures Check Cassette 1 Chamber temperature does not rise more than 2 C i...

Страница 2: ......

Страница 3: ......

Страница 4: ......

Страница 5: ...Calibration Section Check for kinks in the exhaust tubing Check to see if condenser bottle is full of hot water Cycle Fault 9 N A Cycle Fault 10 The cassette temperature has failed to drop to 115 C d...

Страница 6: ......

Страница 7: ......

Страница 8: ...open and close and the solenoid clicks Repair cassette leak and clicking should stop in approximately 24 hours Loud buzzing noise Clean or replace solenoid as needed Noise during drying cycle only So...

Страница 9: ...of cover until it clears internal components then slide cover forward about an inch Gently rotate the cover up and pivot it around the front left corner of the machine When the cover is clear of the m...

Страница 10: ......

Страница 11: ...holes in the new seal with the holes in the lid 6 Install the holes first then install the four corners allowing the sides to hang out Be sure that two indicator squares are visible in each corner 7 S...

Страница 12: ...imately 800 cycles and if maintained properly could last longer Pump Tester Instructions I Disconnect the Teflon boiler inlet tube from the top of the boiler 2 Connect the disconnected end of the inle...

Страница 13: ...he compression nut holding the steam generator boiler outlet tube to the steam generator boiler outlet fitting 8 Remove the two screws from the steam generator boiler bracket The steam generator boile...

Страница 14: ...ting Tighten finger tight 5 Connect the compression nut holding the steam generator boiler outlet tube to the probe bracket inlet fitting Tighten finger tight 6 Connect the compression nut holding the...

Страница 15: ...e bracket Trim away the excess RTV silicone and remove the screws Do Not remove the probe bracket to get at the screws If you have trouble getting at the screws disconnect the gray hose between the ch...

Страница 16: ...the reservoir is filled with good distilled water and the condenser bottle is not full 6 After 15 minutes have passed start an unwrapped cycle while simultaneously activating the pump You may operate...

Страница 17: ...d chamber temperatures The value in the lower right corner of the display is the chamber temperature F9 CALIBRATING C8 2 00 C 135 00 C EXAMPLE 5 To start the steam generator boiler self calibration cy...

Страница 18: ...f the Statim on Press the WRAPPED button again to scroll to the next language Press the UNWRAPPED button to scroll to the previous language When the desired language is displayed press the STOP button...

Страница 19: ...button Press the button repeatedly to select subsequent fields 7 To save changes and return to regular operating mode press the STOP button Diagnosing Water Quality Sensor Problems The water quality...

Страница 20: ......

Страница 21: ......

Страница 22: ......

Страница 23: ......

Страница 24: ......

Страница 25: ...x 7 1 2 Cassette Internal Dimensions 15 x 7 x 1 Reservoir Capacity 4 liters distilled water Approximately 20 cycles Weight Without Water 72 lbs Power Consumption 1300 W Optional Printer Type Thermal...

Страница 26: ...ber and Plastics 40 minutes Heavy Duty Unwrapped 12 minutes Warranty I Year Parts excluding gasket 1 Year SciCan Labor http www sterilizers com statim trouble shooting manual html 26 of 26 3 27 2004 9...