STATIM
To replace the thermal fuse follow these steps:
1. Connect the thermal fuse terminal to the lower power terminal on the steam
generator (boiler).
2. Attach the steam generator (boiler) bracket to the steam generator (boiler)
using the small screw. Route the wire from the fuse assembly between the bracket
and spacing standoffs on the bottom of the steam generator (boiler).
The
fuse must
be placed exactly between the standoffs. DO NOT PINCH OR CRUSH THE
WIRE BETWEEN COMPONENTS.
3. Carefully return and attach the steam generator (boiler) assembly to the chassis
using two screws. Start the screws into the tapped hole but do not tighten the
screws. BE CAREFUL NOT TO STRESS THE THERMOCOUPLE LEADS.
4. Connect the compression nut holding the steam generator (boiler) outlet tube to
the steam generator (boiler) outlet fitting. Tighten finger tight.
5. Connect the compression nut holding the steam generator (boiler) outlet tube to
the probe bracket inlet fitting. Tighten finger tight.
6. Connect the compression nut holding the white Teflon pump tube to the top of
the steam generator (boiler). Tighten finger tight, then continue then continue to
tighten the nut using a 3/8 inch wrench. DO NOT OVERTIGHTEN.
7. Tighten the two screws that attach the steam generator (boiler) assembly to the
chassis.
8. Using a 7/16 wrench, tighten the compression nut holding the steam generator (boiler)
outlet tube to the steam generator (boiler) outlet fitting. Tighten the compression nut holding
the steam generator (boiler) outlet tube to the probe bracket fitting also.
9. Connect the white wire to the terminal on the lower half of the steam generator
(boiler) using the screw with captive lockwasher. If the terminal appears
blackened, clean it using fine grit sandpaper.
http://www.sterilizers.com/statim-trouble-shooting-manual.html (14 of 26) [3/27/2004 9:23:44 PM]