Service Menu Overview
6. Technical Service Menu
Cycle
Count
Diagnostic
Tools
• Error History
• Test Cycle
• Component Test
• View IO Set up
• Verify Mode
• Set debug screen
Cycle
Settings
Printer
Selection
Baud
Rate
Software
Upgrade
Factory
Default
Repeater
Mode
Production
Tools
• Set Drying Time
• Set Regeneration
• Chemical set-up
• Set Wash Time
• Set Rinse Time
• # Extra Rinses
• Set Wash Temperature
• Last Rinse Temp
To access this menu, turn the unit on. There is an "i" in the lower right hand corner of the screen for about
10 seonds. Touch the "i" to get to the menu screen. Touch Service.
Key in the password 7919 and press EN.
Within main service menu, there are eight options.
Cycle Count:
Displays the number of cycles that have been run (complete and incomplete).
Diagnostic Tools:
Offers a submenu of five tools.
Error History: Allows access to the last three errors.
Test Cycle: An abbreviated cycle that uses all pumps and valves.
Cannot be used to process instruments.
Component Test: Allows individual testing of the following components.
Circ. Pump M1 - ON/OFF
All Devices - ON/OFF
Latch L1 - ON/OFF
Salt Valve Y2 - ON/OFF
Dosing Pump 1 M4 - ON/OFF
Dryer Motor + Heater - ON/OFF
Hot W. Valve Y7- ON/OFF
Cold W. Valve Y1 - ON/OFF
Rinse Aid Valve ON/OFF (not used)
Waste Pump M2 - ON/OFF
Page 19
HYDR
IM
C51wd Service Guide