background image

8. Spare Parts  

Item

SCICAN #

KIT DESCRIPTION

NOTE

1

01-107786S

Seal Door, J

2

01-107787S

Inlet Hose Europe, J

one hose / kit

3

01-107789S

Drain Hose, J

one hose / kit

4

01-107790S

Dosing Pump, J

5

01-107791S

Valve Salt Regeneration, J

6

01-107792S

Thermoactuator, J

7

01-107794S

Pump Recirculation Lower Arm, J

8

01-109779S

Pump Top Arm Recirc, S

9

01-107796S

Drain Pump Europe 230V, 50Hz, J

10

01-107799S

2A Fuse, J

five fuses / kit

11

01-107800S

10A Fuse, J

ten fuses / kit

12

01-107801S

Inlet Valve Cold Water, J

13

01-107802S

Switch Full Chamber, J

14

01-107803S

Switch Overflow Chamber, J

15

01-107804S

Lower Wash Arm, J

16

01-107806S

Screen Drain, J

with mesh

17

01-107807S

Filter Drain, J

18

01-107808S

Water Heater, J

19

01-107810S

Switch Pressure Heater, J

20

01-107811S

Door Spring Kit, J

21

01-107812S

Clip Door, J

22

01-107815S

Inlet Valve Hot Water, J

23

01-107938S

Switch Pressure Assy, J

24

01-107975S

Adjustable Feet, J

25

01-108030S

Quick Connect Female, J

26

01-108121S

Cap Quick Disconnect, J

ten caps / kit

27

01-108122S

Drip Tray, J

28

01-108253S

Reservoir

28

01-108305S

Hydrim Water Test Kit, J/K

30

01-108309S

Kickplate, J

31

01-108310S

Mesh Drain, J

only mesh

32

01-108351S

Float Dosing Reservoir, J

33

01-108699S

Tube Chem.

34

01-108700S

Fuse Holder, J

three parts / kit

35

01-108795S

Plastic Trim Edge, J

36

01-108797S

Support Bracket 1-st Pump, J

37

01-108924S

Screw Kit Hydrim C51wd, J

38

01-109143S

Dryer Tubing, J

39

01-109144S

Dryer Fitting, J

For an updated parts list, please refer to myscican.com

Page 33

HYDR

IM

C51wd Service Guide

Содержание HYDRIM C15wd

Страница 1: ...www scican com Service Manual HYDRIM C51wd Service Manual 95 109724 Rev 3 0 HYDRIM C51wd...

Страница 2: ...Inlet Hoses 2 5 Drain Requirements 2 6 Levelling the Hydrim 2 7 Installing Cleaning Solution 2 8 Setting Water Softener 2 9 Installation Test 2 10 Printer USB Setup 3 Routine Maintenance Page 12 3 1 F...

Страница 3: ...including lost profits any commercial loss economic loss or loss arising from personal injury The Hydrim C51wd Instrument Washer Disinfector should only be installed and serviced by a qualified contr...

Страница 4: ...Category II Maximum relative humidity 80 for temperatures up to 31 C 50 for temperatures up to 40 C Operating temperature range 5 C to 40 C Maximum altitude 2000 m Mains supply 10 of nominal Page 3 HY...

Страница 5: ...t be performed on the unit if parts associated with the power main are serviced or replaced Ground Aprotective bonding impedance test ground continuity must be performed on the unit if components of t...

Страница 6: ...drain the dosing reservoir Completely screw in the levelling legs Specify upright heated and insured shipping DESCRIPTION 1 Nose pliers 2 Screwdriver PH1 3 Screwdriver PH2 4 Screwdriver Slot 5 T20 To...

Страница 7: ...ponsible for damage or injury caused by incorrect or missing grounding The Hydrim unit is heavy 35 kg Exercise caution and obtain assistance when moving it The Hydrim is equipped with an air gap anti...

Страница 8: ...ce access and air exhaust from the rear of the unit Do not locate the electrical outlet directly behind the Hydrim as this is where warm moist air from the chamber is exhausted Do not move the Hydrim...

Страница 9: ...hould be attached to the back of the Hydrim unit The washer with the screen goes to the water supply connector Make sure that the inlet valves are free of debris 2 Installation drain hose outlet cold...

Страница 10: ...al codes 2 6 Levelling the Hydrim For the unit to function properly it will need to be correctly levelled To level the unit follow these steps 1 Adjust the legs underneath the unit 2 Use the levelling...

Страница 11: ...pe Power the unit on Touch the i in the lower right hand corner of the sceen Select Technician Enter 7919 and touch EN Select Cycle Settings and then Set Regeneration Using the up and down arrows set...

Страница 12: ...er Selection 3 Select Serial Printer if connecting a printer or USB Flash MSD if connecting the SciCan Data Logger Press the back arrow to return to the Setup Menu 4 In the Setup Menu select Baud rate...

Страница 13: ...the fitting at the hub note fitting is left threaded 3 Remove the upper wash arm 4 Using two hands grasp both ends of the lower wash arm on the underside 5 Pull the lower wash arm upwards 6 Inspect b...

Страница 14: ...air filter 47 dosing reservoir 28 heater 18 lower circulation pump 7 water softener door solenoid 62 dryer duct 39 top arm fitting 65 Tubing top arm fitting 64 door lock level sensor valves hot and c...

Страница 15: ...Figure 16b 6 Phillips flat head machine screws on the inside front of the unit 4 on the top 2 on the left Figure 16c The two screwsontherightsidedonotsecure thetopcoveranddonotneedtobe removed 4 Gras...

Страница 16: ...Edges are sharp 9 Be careful not to damage the bottom pan overflow float and make sure it is in place before reinstalling bottom pan 4 4 Removing the Chemical Bracket 1 Power the unit OFF 2 Remove the...

Страница 17: ...he fascia to the chamber Figure 18c 4 Pull chemical bracket away and swing fasica out as shown Figure 18d 5 Access the controller assembly and disconnect the following connectors from the i o board J6...

Страница 18: ...e 17 HYDRIM C51wd Service Guide J6 J5 J2 J1 J3 J7 Figure 18f Figure 18g screws screws standoffs printer cable 6 Disconnect the signal connector from the bottom of the logic board Figure 18e 7 Unclip t...

Страница 19: ...y B9 Distribution Valve Motor Switch Eschmann Plus only B10 Conductivity Switch Hydrim L110wd only B11 Handpiece water filter pressure switch Hydrim C51wd LCS only B12 Temperature cut off switch B13 D...

Страница 20: ...the menu screen Touch Service Key in the password 7919 and press EN Within main service menu there are eight options Cycle Count Displays the number of cycles that have been run complete and incomple...

Страница 21: ...s I O status when cycle is running Cycle Settings Set Drying Time P1 Regular 10 min choose from 1 to 15 min P2 Heavy Duty 10 min choose from 1 to 15 min P3 HD Disinfection 10 min only value accepted i...

Страница 22: ...Disinfection 50 C choose from 45 to 60 C Last Rinse Temp P1 Regular 60 C choose from 30 to 80 C P2 Heavy Duty 60 C choose from 30 to 80 C P3 HD Disinfection 80 C only value accepted is 80 Printer Sele...

Страница 23: ...l Switch OFF unit is still filling Chamber overflow switch is OFF Door latch is closed Detergent switch is ON there is enough chemical for dosing Chamber pressure switch is ON upper arm has enough pre...

Страница 24: ...nitor the pressure switch during disinfection P3 only If the switch turns OFF the pump is overheating Check and replace the cooling fan and or the upper circulation pump Figure 20 If OK proceed to ste...

Страница 25: ...cket Check the level diaphragm switch assembly and repair or replace If OK proceed to step 5 5 Possible cause is a leak Remove the kick plate and look for fluid in the pan Isolate and repair the sourc...

Страница 26: ...1 Check for a kink or blockage in the drain hose 2 Possible cause is drain pump failure Manually load 3 litres of water into the chamber With the troubleshooting window in place go into the service m...

Страница 27: ...rd Replace logic board 7 2 7 CF 7 Cycle Aborted or Interrupted C51WD only Detection Second temperature sensor readings are out of limits either too low temperature or too high temperature Cause Broken...

Страница 28: ...ds is not achieve during Circulation and heating phase and the pressure switch for the upper wash arm is OFF Cause Defective upper arm pressure switch 7 2 11 CF 12 Handpiece washing failure C51WD only...

Страница 29: ...late Check the connectors or replace the temperature sensor as required 7 2 13 CF 15 Chamber Overflow Detection The overflow switch did not turn off after 30sec of running the drain pump Remedy 1 Chec...

Страница 30: ...tive pressure switch 7 2 15 CF 17 Circulation Pressure Sensor stuck Detection Circulation Pressure sensor stuck on high pressure at Idle or during first phase of the filling stage Cause Circulation Pr...

Страница 31: ...voir malfunction Remove the top cover Remove two small red wires from the connector block and connect them together Figure 28 If the no detergent message disappears and the machine runs the dosing res...

Страница 32: ...ks Replace if necessary Problem Water leaking from the door front of the unit Remedy 1 Make sure unit is level 2 Check adjustment on the door latch Loosen two screws and then slide door latch in or ou...

Страница 33: ...oesn t open Remedy 1 Check the connector from the I O board to the latch 2 Check the solenoid and replace if necessary 3 Check the 24V power supply 4 Check if mechanical link is broken 5 Replace the I...

Страница 34: ...r Drain J 18 01 107808S Water Heater J 19 01 107810S Switch Pressure Heater J 20 01 107811S Door Spring Kit J 21 01 107812S Clip Door J 22 01 107815S Inlet Valve Hot Water J 23 01 107938S Switch Press...

Страница 35: ...S Cooling Fan Hydrim S 51 01 109791S Door Curtain Hydrim S 52 01 109790S Upper SprayArm Hydrim S 53 01 109789S Rope Door Latch C51WD S 54 01 109788S Dual Temperature Sensor C51WD S 55 01 109787S Plug...

Страница 36: ...Recirc S 01 107796S Drain Pump Europe 230V 50Hz J 01 107810S Switch Pressure Heater J 01 107812S Clip Door J 01 107815S Inlet Valve Hot Water J 01 107799S 2AFuse J 01 107800S 10AFuse J 01 107801S Inle...

Страница 37: ...ir 01 108309S Kickplate J 01 109145S Dryer VentAssy J 01 109776S TopArm Fitting Hydrim S 01 109778S Water Pressure Switch C51WD S 01 109777S Tubing TopArm Fitting Hydrim S 01 108351S Float Dosing Rese...

Страница 38: ...01 109787S Plug Test Port C51WD S 01 109788S Dual Temperature Sensor C51WD S 01 109789S Rope Door Latch C51WD S 01 109790S Upper SprayArm Hydrim S 01 109792S Cooling Fan Hydrim S 01 109793S Fascia Do...

Страница 39: ...e 38 HYDRIM C51w Service Guide 8 Spare Parts 01 110888S 01 110855S tubes white_clear_wheel_thing 01 110837S 01 110842S MAM 005 01 110853S 01 110814S 01 110839S 01 110850S 01 110848S 01 110849S 01 1108...

Страница 40: ...itional step Compressed Air upply Acompressed air feed must be connected to the air connection socket Note Maximum air supply pressure 1 0 BAR The air supply regulator should be adjusted to 0 75 0 85...

Страница 41: ...ycle and proceed with sterilization Do not leave the handpieces in the unit overnight or for any extended length of time Using The LCS With Other Accessories The handpieces occupy the upper right quad...

Страница 42: ...ch or damage the two water tubes Disconnect the quick connect Unscrew the cap from the base and rinse with water to remove an debris Unscrew the old filter from the base and discard Remove any visible...

Страница 43: ...ull the old filter away at the right hand side Then pull to the right remove and discard The flexible port on the right of the filter nest is also displaced to the right to provide clearance to instal...

Страница 44: ...LCS feed line assembly then unscrew the 7 screws connecting the LCS to the Hydrim door Then follow the tubes down through the kickplate and bottom pan to the manifold Disconnect the air check valve a...

Отзывы: