7. Troubleshooting
7.2.5 CF 5 Disinfection Phase Failure
Detection:
Chamber temperature dropped below 80˚C during the disinfection phase of the cycle
Remedy:
Possible cause is a water heating problem. Follow the procedure in CF1.
7.2.6 CF 6 Serial Communication Failure
Detection:
Failure to read second temperature sensor within 10 seconds.
Remedy:
Possible causes are a corrupted program or flash memory, or a defective logic board.
Replace logic board.
7.2.7 CF 7 Cycle Aborted or Interrupted(C51WD only)
Detection:
Second temperature sensor readings are out of limits (either too low temperature or too high
temperature).
Cause:
Broken second temperature sensor wire, bad second temperature sensor connection to PCB
7.2.8 CF 8 Secondary Chamber Temperature Sensor Failure
Detection:
Second temperature sensor readings are outside limits (too high or too low).
Remedy:
Possible causes are disconnection or failure of the temperature sensor. Remove the kick plate.
Check the connectors or replace the temperature sensor as required.