2.1 Pre-Installation
The machine must be installed and levelled correctly for the unit to function as described.
All electrical work must be carried out by a qualified electrician and in compliance with all
local and national electrical codes.
The electrical outlet should not be located directly behind the unit. The outlet needs to be
accessible after the unit is installed. The power cord must be routed away from the back
panel and hot water inlet hose.
This appliance must be correctly grounded! The manufacturer cannot be held
responsible for damage or injury caused by incorrect or missing grounding.
•
The Hydrim unit is heavy (35 kg). Exercise caution and obtain assistance when
moving it.
•
The Hydrim is equipped with an air gap / anti-suction device to prevent backflow of
dirty water into the water supply. No other air gap device is necessary.
•
If you need to extend the water inlet and drain hoses,ensure that you use commercial
grade plumbing hose. The maximum length of the drain hose is 3.3 m.
2.2 Tools and Supplies Required to Install the Hydrim
• Slot screwdriver
• Channellocks
Ensure that HIP Cleaning Solution (instrument wash chemical) is available. All other
supplies are included with the Hydrim unit.
Voltage
:
220 - 240 V
Frequency
: 50 Hz
Rated load
:
2 kW
Circuit breaker
:
10 A per phase
2. Installation