11.1 Appendix - Additional Information For The C51wd-LCS Model
This Appendix provides the additional information required for the C51wd-
LCS model which comes equipped with a Lumen Cleaning System (LCS).
Intended Use
The Lumen Cleaning System is intended for cleaning and drying of
instruments with internal channels (e.g. dental handpieces). The system is
not intended to disinfect the channels. The handpieces are attached via
adaptors to the door of the C51wd-LCS and are cleaned and dried during
the normal cycle.
Installation Instructions
The C51wd-LCS, equipped with the Lumen Cleaning System, is installed in
the same way as the C51wd but with the following additional step:
Compressed Air upply
A compressed air feed must be connected to
the air connection socket.
Note: Maximum air supply pressure 1.0 BAR.
The air supply regulator should be adjusted to
0.75 - 0.85 Bar.
The compressed air supply is required for
validated handpiece cleaning and drying.
Instructions for Use
The unit is used and cycles operate in the same
way as the C51wd. ONLY use P2 or P3 cycles
when handpieces are attached.
11.2 Attaching Handpieces To The LCS
The C51wd-LCS comes equipped with
six universal couplings and can
process up to six dental handpieces
per cycle, in addition to standard
instruments. Adaptors that are specific
to handpieces should be fitted to each
universal coupling being used.
CAUTION: When adaptors or
handpieces are fitted to the LCS, the trolley cannot be fully withdrawn. Take
care not to damage adaptors or handpieces when withdrawing the trolley.
air connector
socket