General Information
Intended use
The EMP 4 PRO device has been designed for the transcutaneous stimulation of muscles
and nerves and may not be used for any other purposes.
Safety instructions
Please read the operating instructions carefully before using the EMP 4 PRO.
• The EMP 4 PRO is only to be used for stimulation purposes on one patient at any one time.
• Only use the EMP 4 PRO with original accessories.
• The size of the electrodes should to be not less than 2 cm².
• The electrodes should be positioned in such ways that the current flow will not cross
through the heart.
• Keep the EMP 4 PRO away from water or other liquids. Do not drop the EMP 4 PRO, do
not use it inappropriately or expose it to extreme temperatures or high levels of humidity
(not less than 10 °C or more than 40 °C or a relative humidity of more than 90 %).
• Do not use the EMP 4 PRO if it is not working properly or if it has been damaged in any way.
• Be careful when using the EMP 4 PRO near or on children. Keep the unit away from children.
• Always store the EMP 4 PRO in its case to protect it from damage and dust.
• To avoid reciprocal interference, the EMP 4 PRO should not be operated in the vicinity
of other electronic devices. If this is not possible, the various unit functions should
be closely monitored during operation so that proper use in accordance with the
regulations is ensured.
• The simultaneous connection to the patient of a high frequency surgical unit can lead
to burning under the electrodes or around the probe.
• Operating close (for example 1m) to short wave and microwave units or mobile
telephone systems can cause fluctuations in the output values of the electric current
stimulation unit.
• Do not use the machine during the simultaneous operation of other machines, when
driving or when asleep.
Contra indication
Please check with your physician before using the EMP 4 PRO if one of these
apply to you:
• Patients with electronic implants such as heart pacemakers, defibrillators or pumps
• Patients with cardiac rhythm disorders
• During pregnancy