schwa-medico
EMP 4 PRO
14
No.
USER Program type
Possible parameters
61
TENS continuous mode
• Frequency (1-100 Hz)
• Impulse width (50-250 μs)
• Stimulation time (10-90 min)
62
Frequency modulation
• Lowest frequency: from 1 to 100 Hz
• Highest frequency: from the lowest
fre 10 Hz to 100 Hz
• Impulse width (50-250 μs)
• Stimulation time (10-90 min)
63
EMS excito-motor
stimulation
• Frequency (1-60 Hz)
• Impulse width (50-350 μs)
• Rising ramp time: 2 s
• Plateau working time: 6 s
• Falling ramp time: 1 s
• Pause time: 6 s
• Stimulation time (10-90 min)
64
Endorphin burst
stimulation
• Frequency (1-100 Hz)
• Impulse width (50-200 μs)
• Rising ramp: 0.25 s
• Stimulation time (10-90 min)
65
Massage through
dynamic stimulation
(one direction)
• Frequency (1-60 Hz)
• Impulse width (50-350 μs)
• Rising ramp time: 2 s
• Plateau working time: 1 s
• Falling ramp time: 1 s
• Stimulation time (10-90 min)
66
Massage through
dynamic stimulation
(both directions)
• Frequency (1-60 Hz)
• Impulse width (50-350 μs)
• Rising ramp time: 2 s
• Plateau working time: 1 s
• Falling ramp time: 1 s
• Stimulation time (10-90 min)
67
EMS
agonist/antagonist
stimulation
• Frequency (1-60 Hz)
• Impulse width (50-350 μs)
• Rising ramp time: 2 s
• Plateau working time: 3 s
• Falling ramp time: 1 s
• Stimulation time (10-90 min)
You must first adjust the intensities of channels 1 and 2 and
then channels 3 and 4. The intensities of the active channels
flashes on the screen.
Overview user programs