schwa-medico
EMP 4 PRO
15
No.
USER Program type
Possible parameters
68
2+2 mode
Ch. 1 & 2: EMS
Ch. 3 & 4: TENS
For channels 1 and 2:
• Frequency (half of frequency from
channel 3 and 4)
• Impulse width (50-200 μs)
• Rising ramp time: 3 s
• Plateau working time: 1 s
• Falling ramp time: 3 s
• Pause time: 6 s
• Stimulation time (10-90 min)
For channel 3 and 4:
• Frequency (1-90 Hz)
• Impulse width (50-200 μs)
• Stimulation time (10-90 min)
The pulse width must be the same for
all 4 channels
Setting parameters for user programs
• Select a user program (61 to 68) depending on the desired type of therapy by pressing
the
P
key (see selecting a program).
• Press the key for 3 seconds to enter the setting mode.
• The first parameter to be changed flashes.
• Press the
▲
and
▼
keys of any channel to change the parameters.
• Press again to go to the next parameter.
• Press the
▲
and
▲
keys of any channel to change the parameters.
• Press again to go to the next parameter.
• Press the
▼
and
▲
keys of any channel to change the parameters.
• Press after changing the last parameter to save the changes in the user program.
Press
P
at any time if you want to abort the settings. The changed parameters up to that
point are then not saved.
Overview user programs