Maintenance and cleaning
No special cleaning or care agents are required for the EMP 4 PRO. Clean the unit with a
soft, fluff-free cloth. Please ensure that no moisture permeates the unit. If moisture does
permeate the unit, a technical check must be carried out before re-use.
Classification of the device
In accordance with the law on medical devices, the EMP 4 PRO is classified as being a
Class IIa medical device.
Technical check
We recommend a technical check on the EMP 4 PRO every 24 months.
This includes:
1. Confirming that user instructions have been included in the
accompanying documentation.
2. Checking the equipment for completeness.
3. Visual check:
- for mechanical damage
- for damage to all cables and sockets
4. Functional Safety
- checking the output signals with a load resistance of 1 k
Ω
real
current and voltage
- checking the frequency
- checking the pulse width
These technical checks may only be carried out by suitably qualified persons. The results
must be documented with the date and name of the person carrying out the check.