Operation
handylab pH 11
30
ba75366e04
01/2011
sponds to a
single-point calibration
. For this, the instrument uses the
Nernst slope (-59.2 mV/pH at 25 °C) and the specified asymmetry of the elec-
trode.
8
To continue the
two-point calibration
, thoroughly rinse the
electrode with deionized water.
9
If necessary, enter the temperature of the second buffer solu-
tion with
<
▲
> <
▼
>
.
10
Immerse the pH electrode in the second buffer solution.
11
Press the
<RUN/ENTER>
key. The second pH value appears
on the display.
12
Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with
<
▲
> <
▼
>
.
13
When the measured value is stable, press the
<RUN/ENTER>
key. The value of the slope (mV/pH) appears on the display.
The probe symbol shows the evaluation of the current calibra-
tion.
14
Press the
<RUN/ENTER>
key. The value of the asymmetry
(mV) appears on the display.
AutoCal DIN
AutoCal TEC
6
mV/pH
mol/L
8
Auto Store
RCL
LoBat
CalError
ARng AR
Time
Baud
Day.Month No.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
O
SL
°C
TP
pH
ConCal
Содержание handylab pH 11
Страница 1: ...ba75366d04 01 2011 Operating manual handylab pH 11 pH Meter AR O CAL PH 6 4 2 9 9 2 6 C TP M...
Страница 4: ...List of contents handylab pH 11 4 ba75366e04 01 2011...
Страница 10: ...Overview handylab pH 11 10 ba75366e04 01 2011...
Страница 14: ...Commissioning handylab pH 11 14 ba75366e04 01 2011...
Страница 31: ...handylab pH 11 Operation 31 ba75366e04 01 2011 15 To return to the measuring mode Press the M key...
Страница 40: ...What to do if handylab pH 11 40 ba75366e04 01 2011...
Страница 46: ...Index handylab pH 11 46...
Страница 47: ......