Operation
handylab pH 11
20
ba75366e04
01/2011
Displaying
calibration data
Each time the instrument is switched on, the calibration data are shown on
the display for a short time (see section 4.1 S
WITCHING
ON
THE
MEASURING
IN
-
STRUMENT
). In order to view the calibration data, switch the measuring instru-
ment off and switch it on again.
Calibration
evaluation
After the calibration, the measuring instrument automatically evaluates the
current status of the electrode. The asymmetry and slope are evaluated sep-
arately. The worst evaluation appears on the display.
Preparatory
activities
Display
Asymmetry
[mV]
Slope
[mV/pH]
-15 ... +15
-60.5 ... -58
-20 ... +20
-58 ... -57
-25 ... +25
-61 ... -60.5
or
-57 ... -56
Clean the electrode according
to the electrode operating
manual
-30 ... +30
-62 ... -61
or
-56 ... -50
E3
Eliminate the error according to
chapter 6 W
HAT
TO
DO
IF
...
< -30 or
> 30
< -62 or
> -50
1
Switch on the measuring instrument with
<ON/OFF>
.
2
Connect the pH electrode to the measuring instrument.
3
Keep the buffer solutions ready.
4
Adjust the temperature of the solutions and measure the cur-
rent temperature if the measurement is made without temper-
ature sensor.
Содержание handylab pH 11
Страница 1: ...ba75366d04 01 2011 Operating manual handylab pH 11 pH Meter AR O CAL PH 6 4 2 9 9 2 6 C TP M...
Страница 4: ...List of contents handylab pH 11 4 ba75366e04 01 2011...
Страница 10: ...Overview handylab pH 11 10 ba75366e04 01 2011...
Страница 14: ...Commissioning handylab pH 11 14 ba75366e04 01 2011...
Страница 31: ...handylab pH 11 Operation 31 ba75366e04 01 2011 15 To return to the measuring mode Press the M key...
Страница 40: ...What to do if handylab pH 11 40 ba75366e04 01 2011...
Страница 46: ...Index handylab pH 11 46...
Страница 47: ......