handylab pH 11
Operation
29
ba75366e04
01/2011
4.3.3
ConCal
Single-point
calibration
Use any buffer solution for this rapid method.
The calibration will be the more exact the nearer the pH value of the buffer
solution is to that of the test sample.
Two-point
calibration
For this procedure, use two buffer solutions:
z
pH 7.0 ± 0.5
z
any other buffer solution
Note
Skip the steps 2 and 9 if you use a pH electrode with a temperature sensor.
Note
At this point, the ConCal calibration can be broken off with
<M>
. This corre-
1
Press the
<CAL>
key repeatedly until the display
ASY
and the
function display
ConCal
appears. The sensor symbol displays
the evaluation of the last calibration (or no sensor symbol in the
delivery state or after the measurement parameter has been
reset).
2
If necessary, enter the temperature of the first buffer solution
with
<
▲
> <
▼
>
.
3
Immerse the pH electrode in the first buffer solution (pH 7.0 ±
0.5 in two-point calibration).
4
Press the
<RUN/ENTER>
key.
The measured pH value appears on the display.
5
Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with
<
▲
> <
▼
>
.
6
Press the
<RUN/ENTER>
key. The value of the asymmetry
(mV) and the sensor symbol appear on the display.
7
Press the
<RUN/ENTER>
key.
SLO
appears on the display.
Содержание handylab pH 11
Страница 1: ...ba75366d04 01 2011 Operating manual handylab pH 11 pH Meter AR O CAL PH 6 4 2 9 9 2 6 C TP M...
Страница 4: ...List of contents handylab pH 11 4 ba75366e04 01 2011...
Страница 10: ...Overview handylab pH 11 10 ba75366e04 01 2011...
Страница 14: ...Commissioning handylab pH 11 14 ba75366e04 01 2011...
Страница 31: ...handylab pH 11 Operation 31 ba75366e04 01 2011 15 To return to the measuring mode Press the M key...
Страница 40: ...What to do if handylab pH 11 40 ba75366e04 01 2011...
Страница 46: ...Index handylab pH 11 46...
Страница 47: ......