handylab pH 11
Operation
19
ba75366e04
01/2011
4.3
Calibrating
Why calibrate?
pH electrodes age. This changes the asymmetry (zero point) and slope of the
pH electrode. As a result, an inexact measured value is displayed. Calibration
determines the current values of the asymmetry and slope of the electrode
and stores them in the measuring instrument.
Thus, you should calibrate at regular intervals.
When to calibrate?
z
After connecting another electrode
z
after a voltage interruption, e. g. after changing the batteries
You can select between 3 calibration procedures:
AutoCal TEC
is specially adapted to the SI Analytics technical buffer solutions as a fully au-
tomatic
two-point
or
three-point calibration
. The buffer solutions are auto-
matically recognized by the measuring instrument. Depending on the
instrument setting (see section 4.4 C
ONFIGURATION
), the instrument displays
the relevant buffer nominal value or the current electrode voltage in mV. The
calibration can be terminated after the first buffer solution. This corresponds
to a
single-point calibration
. For this, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the asymmetry of the electrode.
AutoCal DIN
is specially adapted to the permanently programmed buffer solutions in ac-
cordance with DIN19266 as a fully automatic
two-point
or
three-point cali-
bration
. The buffer solutions are automatically recognized by the measuring
instrument. Depending on the instrument setting (see section
4.4 C
ONFIGURATION
), the instrument displays the relevant buffer nominal val-
ue or the current electrode voltage in mV. The calibration can be terminated
after the first buffer solution. This corresponds to a
single-point calibration
.
For this, the instrument uses the Nernst slope (-59.2 mV/pH at 25 °C) and de-
termines the asymmetry of the electrode.
ConCal
This function is a conventional
two-point calibration
using two buffer solu-
tions (pH 7.0 ± 0.5 and any other buffer solution) or a
single-point calibra-
tion
using any buffer solution and is used as a high-speed method.
AutoRead
When calibrating with AutoCal TEC and AutoCal DIN, the
AutoRead
function
is automatically activated.
The current AutoRead measurement can be terminated at any time (accept-
ing the current value) by pressing
<RUN/ENTER>
.
Содержание handylab pH 11
Страница 1: ...ba75366d04 01 2011 Operating manual handylab pH 11 pH Meter AR O CAL PH 6 4 2 9 9 2 6 C TP M...
Страница 4: ...List of contents handylab pH 11 4 ba75366e04 01 2011...
Страница 10: ...Overview handylab pH 11 10 ba75366e04 01 2011...
Страница 14: ...Commissioning handylab pH 11 14 ba75366e04 01 2011...
Страница 31: ...handylab pH 11 Operation 31 ba75366e04 01 2011 15 To return to the measuring mode Press the M key...
Страница 40: ...What to do if handylab pH 11 40 ba75366e04 01 2011...
Страница 46: ...Index handylab pH 11 46...
Страница 47: ......