handylab pH 11
Operation
17
ba75366e04
01/2011
4.2.2
Measuring the pH value
Setting the
resolution
The measuring instrument shows the pH value with a resolution of 0.01 or
0.001. (Default setting is 0.01). To change the resolution, press the
<M>
key
while pressing the
<RUN/ENTER>
key.
AutoRead
(drift control)
The AutoRead function (drift control) checks the stability of the measurement
signal. The stability has a considerable impact on the reproducibility of the
measured values.
For identical measurement conditions, the following criteria apply:
1
Perform the preparatory activities according to section 4.2.1.
2
Immerse the pH electrode in the test sample.
3
Press the
<M>
key until
pH
appears on the status display. The
pH value appears on the display.
mV/pH
mol/L
8
Auto Store
TempError
RCL
LoBat
AutoCal DIN
CalError
ARng AR
AutoCal TEC ConCal
Time
Baud
Day.Month No.
Ident
Year
1
U
%
°F
S
mg/L
% ppm
ISE
8
4
2
2
99
6
°C
TP
pH
Reproducibility
Response time
Better than 0.02
> 30 seconds
1
Call up the pH measuring mode with
<M>
and
<
▲
> <
▼
>
.
2
Activate the AutoRead function with
<AR>
.
The current measured value is frozen (hold function).
3
Start AutoRead with
<RUN/ENTER>
.
AR
flashes until a stable measured value is reached.
This measured value is transmitted to the interface.
4
If necessary, start the next AutoRead measurement with
<RUN/ENTER>
.
5
To terminate the AutoRead function: Press the
<AR>
key.
Содержание handylab pH 11
Страница 1: ...ba75366d04 01 2011 Operating manual handylab pH 11 pH Meter AR O CAL PH 6 4 2 9 9 2 6 C TP M...
Страница 4: ...List of contents handylab pH 11 4 ba75366e04 01 2011...
Страница 10: ...Overview handylab pH 11 10 ba75366e04 01 2011...
Страница 14: ...Commissioning handylab pH 11 14 ba75366e04 01 2011...
Страница 31: ...handylab pH 11 Operation 31 ba75366e04 01 2011 15 To return to the measuring mode Press the M key...
Страница 40: ...What to do if handylab pH 11 40 ba75366e04 01 2011...
Страница 46: ...Index handylab pH 11 46...
Страница 47: ......