Operation
handylab pH 11
18
ba75366e04
01/2011
Note
The current AutoRead measurement can be terminated at any time (accept-
ing the current value) by pressing
<RUN/ENTER>
.
4.2.3
Measuring the ORP voltage
In conjunction with an ORP electrode, e.g. BlueLine 31Rx, the measuring in-
strument can measure the ORP voltage (U) of a solution.
Note
ORP electrodes are not calibrated. However, you can check ORP electrodes
using a test solution.
1
Perform the preparatory activities according to section 4.2.1.
2
Immerse the ORP electrode in the test sample.
3
Press the
<M>
key until U appears on the status display.
The ORP voltage (mV) of the test sample appears on the dis-
play.
4
Wait for a stable measured value.