handylab pH 11
Operation
25
ba75366e04
01/2011
4.3.2
AutoCal DIN
For this procedure, use two or three different standard buffer solutions ac-
cording to DIN 19266 in ascending or descending order (type A, C, D or F
with pH values at 25 °C of: 1.679 / 4.006 / 6.865 / 9.180).
Note
The buffer solutions are automatically recognized by the measuring instru-
ment. Depending on the instrument setting (see section 4.4 C
ONFIGURATION
),
the instrument displays the relevant buffer nominal value or the current elec-
trode voltage in mV.
Note
Skip the steps 2, 7 and 14 if you use a temperature sensor.
1
Press the
<CAL>
key repeatedly until the display
Cd1
and the
function display
AutoCal DIN
appear. The sensor symbol dis-
plays the evaluation of the last calibration (or no sensor symbol
in the delivery state or after the measurement parameter has
been reset).
2
If necessary, enter the temperature of the buffer solution with
<
▲
> <
▼
>
.
3
Immerse the pH electrode in the first buffer solution.
4
Press the
<RUN/ENTER>
key.
The
AR
display indicator flashes.
The electrode voltage (mV) or the buffer nominal value ap-
pears on the display. Example:
AutoCal TEC
6
mV/pH
mol/L
8
Auto Store
RCL
LoBat
CalError
ARng AR
ConCal
Time
Baud
Day.Month No.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
1
C d
°C
TP
pH
AutoCal DIN
Содержание handylab pH 11
Страница 1: ...ba75366d04 01 2011 Operating manual handylab pH 11 pH Meter AR O CAL PH 6 4 2 9 9 2 6 C TP M...
Страница 4: ...List of contents handylab pH 11 4 ba75366e04 01 2011...
Страница 10: ...Overview handylab pH 11 10 ba75366e04 01 2011...
Страница 14: ...Commissioning handylab pH 11 14 ba75366e04 01 2011...
Страница 31: ...handylab pH 11 Operation 31 ba75366e04 01 2011 15 To return to the measuring mode Press the M key...
Страница 40: ...What to do if handylab pH 11 40 ba75366e04 01 2011...
Страница 46: ...Index handylab pH 11 46...
Страница 47: ......