Page 7.3
Chapter 7
Technical Data
SCHILLER AT-2plus 6-Channel ECG Unit
SERVICE HANDBOOK Issue d May 2002
Technical Data
Myogram Filter (muscle tremor filter)
25 Hz or 35 Hz, programmable (not active on averaged
waveform). The stored ECGs can be printed with or without
filter.
Line Frequency Filter:
Distortion-free suppression of superimposed 50 or 60 Hz
sinusoidal interferences by means of an adaptive digital filter.
Patient Input:
Fully floating and isolated, defibrillation protected.
Safety Standard:
CF according to IEC and complying with the following
RL 93/42/EEC
EN 60601-1:1990
IEC 601-1
IEC 601-2-25:1993
pr EN 1441:1994
EMC:
CISPR 11: 1985, EN 55011: 1992
IEC 801-2: 1991
IEC 801-3: 1984
IEC 801-4: 1988
IEC 801-5:
Safety Class:
I according to IEC 601-1 (with internal power supply)
IIa according to RL 93/42/EEC, CE-0123
This device is not designed for outdoor use (IP 20)
Environmental Conditions:
Temperature, Operating: 10
O
to 40
O
C
Temperature, Storage:
-10
O
to 50
O
C
Relative humidity:
25 to 95% (non condensing)
Atmospheric pressure:
700 to 1060 hPa
Control Panel:
Rubber keys
Available Configurations
The CARDIOVIT AT-2
plus
is available in several versions:
Standard Version:
Unit with ECG recording and printout capabilities.
Version C:
Unit with additional ECG Interpretation program (including
measurements).
AT-2
plus
Memory:
Unit with FLASH EPROM memory and RS-232
communication interface.
Содержание AT-2plus
Страница 43: ...Page 2 8 SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Chapter 2 Functional Overview ...
Страница 81: ...Page 6 4 SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Chapter 6 Spare Parts ...
Страница 85: ...Page 7 4 SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Chapter 7 Technical Data ...
Страница 93: ...SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Page 8 8 Chapter 8 Glossary ...