Page 1.13
Chapter 1
Operating Elements
SCHILLER AT-2plus 6-Channel ECG Unit
SERVICE HANDBOOK Issue d May 2002
Settings (cont.)
The defined formats and settings for your unit can be checked as follows:
ALT - 0 - 1 - 1
A printout of the defined settings will be produced and gives the following information, depending
on the installed software:
Unit designation
Software version, Software option installed (C = Interpretation)
and interpretation version
Serial number
Serial number of the unit
Leads
Standard (S) or Cabrera (C)
ECG Format
Long (ooo), Short (o) or Suppressed (-)
MECG
Average cycles as defined in auto ECG recording setup (e.g. 4 *
3 (25 mm/s) + 2)
Measurements
Enabled (+) or Suppressed (-)
Marks
Enabled (+) or Suppressed (-)
I
nterpretation
Enabled (+) or Suppressed (-)
Selected Rhythm leads
Leads selected for R1, R2 resp.
Automatic Centering
Enabled (+) or Suppressed (-)
Printout of signals
Sequential or Simultaneous
Baseline Filter
0.05, 0.15 or 0.30 Hz
Mains Filter
50, 60 Hz or OFF (-)
Myogram Filter
25 or 35 Hz, ON (+) or OFF (-)