Page 4.15
Chapter 4
Module Removal and Replacement
SCHILLER AT-2plus 6-Channel ECG Unit
SERVICE HANDBOOK Issue d May 2002
LCD screen board
Prerequisite
•
The Warnings and Cautions at the beginning of the Chapter must be observed.
•
The Top Assembly must be removed as detailed previously. All external cable assemblies
must be disconnected.
Tools
•
Cross-bladed screwdriver
Part Number
The part number for the LCD screen board is given in Chapter 6.
CAUTION
THE LCD BOARD CONTENT IS STATIC SENSITIVE; OBSERVE ANTISTATIC
PRECAUTIONS
LCD board Removal
1. Unscrew the four retaining screws securing the LCD board in position.
2. Remove the board.
LCD board Replacement
To replace the LCD board proceed as follows:
1. Position the board and secure at the 4 fixing points
2. Connect the following:
•
LCD high-voltage backlight (dual-wire) and signal cable assemblies
•
keyboard cable assembly
3. Replace the top assembly as explained earlier in this Chapter.
Содержание AT-2plus
Страница 43: ...Page 2 8 SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Chapter 2 Functional Overview ...
Страница 81: ...Page 6 4 SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Chapter 6 Spare Parts ...
Страница 85: ...Page 7 4 SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Chapter 7 Technical Data ...
Страница 93: ...SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Page 8 8 Chapter 8 Glossary ...