SCHILLER AT-2plus 6-Channel ECG Unit
SERVICE HANDBOOK Issue d May 2002
Chapter 3
Fault Diagnosis
Memory and Data Transmission Check
To perform these tests, the following equipment is needed:
•
A personal computer (PC) with the SEMA communication program (SEMACOMM)
installed.
•
An RS-232 cable assembly, Art.No. 2.310 159 for connecting the RS-232 interface on the
AT-2plus with the COM port of the PC.
•
A patient simulator attached to the patient connector on the AT-2plus.
Test Procedure
Command
Screen Display
Display Memory files
ALT - M
Memory ECG list
Erase Memory files
Press and hold ALT and press E
ERASE ALL ?
YES / NO
Confirm
Select YES and press ENTER
"ERASING" flashing
Set Auto save ON
ALT - 0 - 5 - 1
Set Auto erase ON
ALT - 0 - 5 - 3
Enter patient data
Press data/ECG button
Patient data
Start AUTO mode ECG
Press AUTO button
ECG
Check Memory files
ALT - M
New file in memory
Connect RS-232 cable between
AT-2plus and PC
Check settings SEMA
Config
Baud, n, 2, 100 ms
Set baud rate AT-2plus
ALT - 0 - 9 - x (see Page 1.25)
Confirm baud rate setting
Press STOP
Set Line transmission
ALT - 0 - 9 - 2 - 1
Enter memory menue
ALT - M
Memory file(s) displayed.
Transmit Memory file(s)
Press and hold ALT and press T
"TRANSMITTING" flashing
Stop transmission
Press and hold ALT and press Q
Check that file has benn
transferred to PC
Directory Semacomm \ sdrecs
New file present.
Check that file has disappeared in
AT-2plus
ALT - M
Memory ECG list empty
Back to measurement mode
Press Data/ECG button
ECG traces
Содержание AT-2plus
Страница 43: ...Page 2 8 SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Chapter 2 Functional Overview ...
Страница 81: ...Page 6 4 SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Chapter 6 Spare Parts ...
Страница 85: ...Page 7 4 SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Chapter 7 Technical Data ...
Страница 93: ...SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Page 8 8 Chapter 8 Glossary ...