Page 5.9
Chapter 5
Adjustments
SCHILLER AT-2plus 6-Channel ECG Unit
SERVICE HANDBOOK Issue d May 2002
ECG Amplifier reference voltage
•
Switch the unit on and measure the voltage difference between the +2V reference and the -
2V reference on pins 1 and 7 of operational amplifier U5. Adjust trimmer VR2 to achieve a
voltage difference of 4000 mV ±2mV.
•
Obtain a printout of the test screen by pressing
ALT - 0 - 3 - any number 0...6 + P
Ensure that the Uref+ and the Uref- measurements are both 2000mV ± 20mV. Ensure that the
Udiff reading is 4000mV . Adjust VR3 to achieve a Udif reading of 20mV.
•
Reassemble the unit as detailed in Chapter 4. Re-check the voltage by again obtaining a
service printout.