Page 1.31
Chapter 1
Operating Elements
SCHILLER AT-2plus 6-Channel ECG Unit
SERVICE HANDBOOK Issue d May 2002
Replacing the Recording Paper
The recording paper must be replaced as soon as the end of the paper is indicated by a red stripe
on the lower edge. After the indication first appears, there are about 8 pages left. However, we
recommend that the paper be replaced immediately. If no paper is left, the printing process is
interrupted and a warning is given on the screen. To replace the paper proceed as follows:
•
Use both hands and place fingers under the retaining bar. Pull directly upwards. The paper
tray cover releases.
•
Withdraw the cover from the unit. DO NOT FORCE, THE PAPER TRAY COVER RUNS
FREELY OVER THE DEDICATED RUNNERS.
•
Remove any remaining paper from the paper tray.
•
Place a new paper pack into the paper tray with the printed (grid) side facing upwards.
•
Place the beginning of the paper over the black paper roller on the paper tray cover.
•
Return the paper tray cover in position and press firmly until secure.
•
Press the STOP key to transport the paper to the start position.
Note:
SCHILLER can only guarantee perfect printouts when SCHILLER original chart paper
or chart paper of the same quality is used.
Содержание AT-2plus
Страница 43: ...Page 2 8 SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Chapter 2 Functional Overview ...
Страница 81: ...Page 6 4 SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Chapter 6 Spare Parts ...
Страница 85: ...Page 7 4 SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Chapter 7 Technical Data ...
Страница 93: ...SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Page 8 8 Chapter 8 Glossary ...