Page 4.5
Chapter 4
Module Removal and Replacement
SCHILLER AT-2plus 6-Channel ECG Unit
SERVICE HANDBOOK Issue d May 2002
Physical Overview
The AT-2plus unit is enclosed in a two part, medical standard, moulded plastic case.
The top part contains the keyboard and the LCD screen with the base section containing all the
electronics of the unit, the RS-232 interface, the thermal printer, the paper tray, the battery and
mains transformer.
The electronics of the unit are contained on two printed circuit boards, the main board (MK 14-
10 ) secured on spacers moulded in the base section, and the RS-232 serial interface board (MK
14-11) mounted on spacers in the base assembly.
The battery is secured in position in a moulded recess and the mains transformer is secured on
spacers above the printed circuit board.
The thermal printer is mounted on a paper tray/thermal printer assembly which is secured in the
base section complete.
Test Equipment, Tools, and Accessories
The following list details the tools, test equipment and accessories required to carry out all
functional tests, calibration procedures and adjustments that can be carried out on the AT-2plus.
The test equipment given here is general. If specific recommendation for test equipment is required,
please contact the SCHILLER service department.
•
Digital Voltmeter
•
Selection of cross-bladed, posi-drive and flat-bladed screwdrivers
•
Cleaning agent such as Trichlorethylene
•
Selection of spanners
•
Double-sided tape
•
ECG Patient Simulator
Содержание AT-2plus
Страница 43: ...Page 2 8 SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Chapter 2 Functional Overview ...
Страница 81: ...Page 6 4 SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Chapter 6 Spare Parts ...
Страница 85: ...Page 7 4 SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Chapter 7 Technical Data ...
Страница 93: ...SCHILLER AT 2plus 6 Channel ECG Unit SERVICE HANDBOOK Issue d May 2002 Page 8 8 Chapter 8 Glossary ...