Set RTC Month
: Use the controls to modify the months field, then press ENTER to make the
selection. Numbers range from 01 through 12 for January through December, respectively.
•
Set RTC Date
: Use the controls to modify the date field, then press ENTER to make the selection.
Numbers range from 01 through 31.
•
Set RTC Year
: Use the controls to modify the year field, then press ENTER to make the selection.
Numbers range from 01 through 99.
•
Set RTC Trim
: A Trim value is used to calibrate the oscillator to improve its accuracy. It represents
a signed oscillator correction which is added once every minute. Calibration is performed by
measuring the averaged oscillator output frequency at Test Point TP32 (CLKOUT) which is
configured to be 4.096 kHz. The Trim value to be entered is computed as:
((F
ideal
–
F
measured
) x (32768 / F
ideal
) x 60) / 2
•
Use the controls to modify the signed value, then press ENTER to make the selection. Numbers
range from -255 t255.
8.13
About
Select this menu option by pressing the Enter Key while the display appears as shown in Figure 38.
Figure 38: ABOUT Menu Option
This menu offers a means to present configuration and version control information. Use the controls
to rotate through the various information displays. Pressing the Enter Key at any time will return
control to the main menu.
8.13.1
Procedure
The procedure for navigating the ABOUT submenu follows the control flow diagram depicted in Figure
39.
ABOUT