Figure 29: TARE ADJUST & TARE SECURED Menu Option
This menu provides a means to “zero
-
out” any component of weight or volume attributed
to the
container used to hold a chemical substance whose net weight or net volume is to be measured. It
represents a negative offset value which, when added to the gross measurement, allows a net value to
be displayed. One of two methods can be used to adjust for tare. First, with nothing on the scale, the
user can enter a known tare value in selectable weight or volume units. Second, with an empty container
of unknown weight placed on the scale, the user can adjust the tare value until the net measurement
display reads zero.
8.9.1
Procedure
The procedures for navigating the TARE ADJUST and TARE SECURED submenus follow the control
flow diagrams as depicted in Figure 30 and Figure 31, respectively.
Figure 30: TARE ADJUST Control Flow Diagram
Display Menu if Secure Tare is
“
DISABLED
”
Display Menu if Secure Tare is
“
ENABLED
”
TARE ADJUST
TARE SECURED
TARE ADJUST
Control Flow if Secure Tare is
“DISABLED”
Path taken when ENTER Key pressed.
Path taken when ARROW Key pressed.
Changes Value in
Selected Unit
EXIT
1: TARE ADJUST
Unit? Kilogram
1: TARE ADJUST
Unit? Pound
1: TARE ADJUST
Unit? Gallon
1: TARE ADJUST
Unit? Liter
1: TARE ADJUST
0.0 lb