Positive Voltage Output
–
Voltage Return
Table 6: Voltage Output Interface
Figure 7: Analog Output Board - V
4.6.1
Features:
•
Voltage Out Interface
: A 2-position terminal-block connector to the voltage output circuit.
•
Output Voltage Range Select
: A jumper field that supports user selection of a voltage output
range of either 0-5V or 0-10V. To select a 0-5V voltage output range, install a jumper-shunt on
Jumper-Field P3. Remove any jumper shunt to select a 0-10V voltage output range.
•
Flash Memory
: 2k-bit of non-volatile memory used to store calibration and configuration data.
•
16-bit DAC
: A voltage-output, Digital-to-Analog Converter, whose output feeds a gain stage,
voltage-output operational-amplifier.
•
Voltage Output Op-Amp
: An operational-amplifier that provides a voltage gain stage.
•
Board ID Select
: A jumper-field used to identify the configuration of the voltage range selection
to the system processor. Its configuration should match that of the Output Voltage Range Select
jumper-field to enable accurate menu option display messages.
-
V +
Voltage
Out
Interface
0-5 V or 0-10 V
Output Voltage
Range Select
Flash Memory
16-bit DAC
Voltage Output
Op-Amp
Board ID
Select