64
SCAIME Manual : NU-eNod4B-ETH-E-236708-D.docx
Stability criterion
None, 0.25 0.5, 1, 2d
Defines a stability interval, related to scale interval.
Zero at Startup
[Yes], [No]
Enable the Automatic Zeroing at power-on
Zero Tracking
[Yes], [No]
Enable the Zero tracking function
Legal for trade
Switch
(Available depending
on version)
[Yes], [No]
Enable the legal for trade mode.
The prerequisite conditions for the activation of legal for trade mode
are:
-
The weight unit must be one of the following: mg, g, kg, t, ct,
µg, oz. (possibility to put spaces after unit)
-
The stability criterion must be 0.25 d.
-
Filter settling time must be less than 1 second in transmitter
mode.
-
The quotient (Maximum capacity / Scale interval) must be less
or equal to 6000.
-
On eNod4-C, zero tracking device must not be activated
together with dynamic zero tracking device.
On eNod4-T, C and D this mode has an effect on user calibration. Thus,
this mode must be activated before processing any calibration.
Save Tare in FRAM
[Yes], [No]
It allows saving tare in non-volatile memory (FRAM).
Save Zero in FRAM
[Yes], [No]
It allows saving zero in non-volatile memory (FRAM).
Audit trail (Available depending on version)