18
TO-BA-e-2020
Note:
When removing the transducer, a thin film of coupling agent is occasionally
drawn between the transducer and the surface. If this is the case, the measurement is
made through this film of coupling agent, resulting in a reading that is greater or less
than it should be. This effect is obvious when one thickness value is observed while
the transducer is being used and another value is observed after the transducer is
removed. In addition, measurements through very thick resist films or coatings can
result in the resist film or coating being measured rather than the actual material. It is
the sole responsibility of the user of the instrument to use the instrument correctly and
to detect such effects.
5.6 Set View Mode
Three view modes can be selected to display the measured value:
Normal
Mode, Scan
Mode
and Diff
Mode
.
Normal Mode
[
Normal Mode
]. As shown in the figure on the right, the last thickness
reading is displayed.
Scan Mode
[
Scan Mode
]. In addition to the last thickness value measured, the
minimum and maximum thickness values are also displayed during the measurement.
By pressing ,
the minimum and maximum values are reset.