TO-BA-e-2020
19
Diff Mode
. Both the last thickness reading and the differential thickness value are
displayed (of the absolute thickness value and the nominal thickness value).
Although the gauge is excellent for single point measurements, it is sometimes
advisable to examine a larger area and search for the point with the smallest thickness.
The gauge has a "Scan
Mode
" function that enables it.
In
Normal Mode
, the meter takes and displays ten measurements per second, which
is sufficient for single measurements. However, in Scan
Mode,
the Meter takes more
than ten measurements per second and displays the readings during scanning. While
the transducer is in contact with the material to be measured, the gauge tracks the
minimum and maximum values determined. The transducer can be "scrubbed" over a
surface, and all short signal interruptions are ignored.
5.7 Set normal thickness
In differential measuring mode, the nominal thickness of the test piece must be set.
The setting procedure is as follows:
Press the F1 / F2 key to move the highlighted cursor. Press the arrow keys to increase
/ decrease the values.
Press the key
or the F3 key to confirm the setting.
Press
key to cancel the change and exit the mode.