Two-point measuring method: the probe is measured twice at the same point of
the test sample, whereby the separation plane of the probe remains 90° for two
measurements. The smaller value should be the thickness at the respective
point.
•
Multi-point measurement method: several measurements are made in a circle
of 30 mm diameter, the thickness value of the tested part being the minimum
value.
•
Continuous measurement method: a continuous measurement is carried out
along the specified line at a distance of less than 5 mm using single-point
measurement, the thickness value of the tested part being the minimum value.
6.2 Wall measurement
During the measurement, the probe parting line can run along the pipe axis or the
vertical pipe axis. For larger pipe diameters, the measurement must be carried out on
the vertical axis. For smaller pipe diameters, the measurement shall be carried out in
both directions, with the thickness value being the minimum value.
7
Service
If the device shows other unusual deviations, no permanently mounted parts may be
disassembled or repaired. In this case, contact our customer service department by e-
mail or telephone to initiate the appropriate service procedure.
8
Transport and storage
Keep the device away from vibrations, strong magnetic fields, corrosive media, dirt and
dust. Store the device at normal temperature.