PR 5610 Installation Manual
Commissioning
Sartorius
95
4.8.3.5 Tare key
Select the tare key function [enabled] or [disabled] with
or
; confirm with
.
4.8.3.6 Set zero key
Select the set zero key function [enabled] or [disabled]
with
or
; confirm with
.
4.8.3.7 Quit in mainlevel
If the instrument is configured via terminal or PC the Q command can be used to terminate the communication.
It can be reactivated with [Setup]-[Serial Ports]-[Operator device at]-[Built-in RS232].
Select with
or
whether the communication can be
terminated with [enabled] or [disabled]; confirm with
.
4.8.3.8 Reset on stop + exit
After pressing the keys Stop and Exit simultaneously the boot menu appears (see chapter 4.3.2)
Select whether the keys need to be pressed [1] sec. or [5] sec. or
are without function [disabled] with
or
; confirm with
.
4.8.3.9 S88.01 interface
The S88.01 interface has to be activated if application software packages (e.g. Truckline, InBatch) are used:
Select S88.01 interface activated [on] or not activated [off] with
or
; confirm with
.
4.8.3.10 Software download
To prevent from the risk that a new application can be loaded and started via network (e.g. with the PR 1750
tool), the download function can be protected and only released via action on the keypad (see chapter 4.8.12).
Select the function software download [protected by setup]
or [always enabled] with
or
; confirm with
.