Commissioning with Terminal
PR 5610 Installation Manual
112 Sartorius
5.2.2 Set
time
Configuration Mo 05.July 2004 13:06:16
D
Set date
T
Set time
......
Enter new time [13:06:16]:
•
Press key [T]
•
The following message is displayed:
Enter new time [13: 06:16]:
•
Enter the new time e.g.:
13: 07: 00
•
Press key [
↵
], to remove the display.
•
Check against the header (top right)
5.2.3 Serial port setup
Calling up serial port setup mask
Change serial port parameter
is done from sub-menu
Configuration
by
pressing key [C].
PR1713 Rel. 03.12 20.05.2004 09:05
Serial port setup Mo 05.July 2004 09:31:51
Operator device: Builtin RS232 (protected)
Printer device : none
Remote device : none
Builtin: CON RS232 (locked by operating)
Slot 1: -no serial-
-no serial-
Slot 2:TTY3 RS485
........TTY4 RS232
Slot 3: -no serial-
-no serial-
PR1713 Rel. 03.12 20.05.2004 09:05
Slot 2 RS485 (TTY3) Mo 05.July 2004 09:31:51
Port :Slot 2 RS232 (TTY3)
Protocol :XON/XOFF
Baudrate : 9600
Bits : 8
Parity :None
Stopbits : 1
Devicetyp:Ansi
Echo :Enabled
⇒
When calling up this mask, the communication is disabled.
In the first mask the serial port is the selected, in the second mask the interface parameter are set. The possible
editing functions are displayed below.
•
Save configuration
Exit and save
If changes were made, the system asks e.g.:
Save Slot 2 RS485 ? (Y/N) Press [Y] or [N]
⇒
The changes are effective only after
Exit
!