![Sartorius PR 5610 Скачать руководство пользователя страница 156](http://html1.mh-extra.com/html/sartorius/pr-5610/pr-5610_installation-manual_1199539156.webp)
PR 5610 Installation Manual
Technical Data
Sartorius
153
6.3 Environmental
effects
6.3.1 Power voltage sags
Power voltage sags
>10 ms at 100% sag, without instrument reaction
>20 ms at 50% sag, " "
"
"
6.3.2 Environmental
conditions
Temperature range
Ambient temperature operation
-10... +40 °C
Switch-on temperature
0... +40 °C
Storage/ transport
-40... +70 °C
Humidity
< 95 %, without condensation, (acc. to IEC 68-2)
Protection type
IP 65
Vibration
to IEC 68-2-6, test Fc
6.3.3 Electromagnetic
immunity
Electrostatic discharge
min. EN 61000-4-2, level 3
Interference (supply and data cables)
min. EN 61000-4-4, level 2
Electromagnetic fields
min. EN 61000-4-3, level 2
6.3.4 RF interference suppression
Electromagnetic emission
acc. to EN55011 group 1, limit value class B