![Sartorius MA37 Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/sartorius/ma37/ma37_operating-instructions-manual_1199627062.webp)
62
Specifications
Operating Instructions Moisture Analyzer MA37
Specifications
Dryer Functions
MA37-1
Temperature range and settings
40°C–160°C, in increments of 1°C; stand-by temperature selectable from 40–100°C
Sample heating
Infrared heating using a metal tube heater
Heating programs
Standard drying, gentle drying
Switch-off criterion
Optional:
– Fully automatic
– Semi-automatic mg (1–50 mg/5–300 s)
– Semi-automatic % (0 .1–5 .0%/5–300 s)
– Time (2:00–99 .59 min .)
– Manual
Displayed result
Moisture in %M and g; dry weight in %S and g; ATRO in %M/S
Net weight, approx .
6 .3 kg
Weighing Functions
Max . weighing capacity
70 g
Repeatability, average
For initial sample weight approx . > 1 g: ± 0 .2%
For initial sample weight approx . > 5 g: ± 0 .05%
Readability
1 mg; 0 .01%
Typical sample quantity
5 to 15 g
External calibration weight
50 g (E2)
Disposable sample pan dimensions
d
90 mm
Interface
Mini USB: Automatic detection of Sartorius printers YDP30 and YDP40,
direct data transfer to Microsoft
®
Windows programs without additional software,
programmable data output interval
Power Supply
Input supply voltage
100–120 V AC and 210–240 V AC (automatic voltage detection), ±10%, 50/60 Hz
Protection class and standards
Protection class I as per EN 61010-1/IEC 61010-1; up to 3000 m above sea level; IP20 as per
EN 60529/IEC 60529
Power supply
Using country-specific power cable from Sartorius only
Power consumption
Max . 640 W (typically)
Ambient Conditions
The specifications are valid for the following ambient conditions:
Environment
For indoor use only
Ambient temperature
+10°C to +30°C
Operating temperature
Operation guaranteed b5°C and +40°C
Storage and shipping
–10°C to +60°C
Height
Up to 3000 m above sea level
Relative humidity
15% to 80% for temperatures up to 31°C non-condensing,
decreasing linearly up to 50% relative humidity at 40°C and 20% at 50°C
Safety of Electrical Equipment
As per EN 61010-1/IEC 61010-1
Safety requirements for electrical equipment for measurement, control, and laboratory use —
Part 1: General requirements
Electromagnetic Compatibility
As per EN 61326-1/IEC 61326-1
Electrical equipment for measurement, control, and laboratory use — EMC requirements —
Part 1: General requirements
Defined immunity to interference:
Suitable for use in industrial areas
Limitation of emissions:
Class B [suitable for use in residential areas and areas that are connected to a low voltage
network that (also) supplies residential buildings] . Therefore, the
analyzer can be used in both areas .
Содержание MA37
Страница 65: ...65 Declaration of Conformity Operating Instructions Moisture Analyzer MA37 Declaration of Conformity ...
Страница 66: ...66 Certificate of Compliance Operating Instructions Moisture Analyzer MA37 Certificate of Compliance ...
Страница 67: ...67 Certificate of Compliance Operating Instructions Moisture Analyzer MA37 ...
Страница 68: ...68 Operating Instructions Moisture Analyzer MA37 Certificate of Compliance ...