![Sartorius MA37 Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/sartorius/ma37/ma37_operating-instructions-manual_1199627046.webp)
46
ISO/GLP-compliant Printout
Operating Instructions Moisture Analyzer MA37
ISO/GLP-compliant Printout
Features
You can have the device information, device ID, and the current date printed
before (GLP header) and after (GLP footer) the values from the weighing series .
The following data is printed:
GLP header:
– Date, time at the start of a weighing series
– Manufacturer
– Model
– Model serial number
– Software version number of the device (BAC)
– Software version number of the display (APC)
– Software version number of the heating module (DOC)
– Device ID with max . 14 characters (if activated under system settings)
GLP footer:
– Field for signature
Configuration
To print out the ISO/GLP report, define the following system settings
(see Chapter “System Settings”, section “Printout”):
t
Activate ISO/GLP-compliant printouts:
In the
Printout
menu under
Printout and Report Format
, activate the
With GLP and method parameters
option .
Operation
t
Print out the report after moisture analysis: Select
a
.
Содержание MA37
Страница 65: ...65 Declaration of Conformity Operating Instructions Moisture Analyzer MA37 Declaration of Conformity ...
Страница 66: ...66 Certificate of Compliance Operating Instructions Moisture Analyzer MA37 Certificate of Compliance ...
Страница 67: ...67 Certificate of Compliance Operating Instructions Moisture Analyzer MA37 ...
Страница 68: ...68 Operating Instructions Moisture Analyzer MA37 Certificate of Compliance ...