Calibration and Adjustment
Operating Instructions Moisture Analyzer MA37
Settings
Language
Date and time
Device information
Calibration/Adjustment
Printout
y
The
Settings
menu appears .
t
Select
Calibration/Adjustment
.
Calibration/Adjustment
MA performance test
Weighing system adjustment
Calibration report
t
Select
Weighing system adjustment
.
t
Select
CAL-Extern
to calibrate the analyzer using an external calibration
weight .
y
The
Calibration/Adjustment
window appears .
t
Close the hood and select
l
to confirm .
t
The
CAL-Extern
window appears .
t
If necessary, please zero the integrated balance .
y
A message appears on the display, prompting you to place the calibration
weight (50 g) on the analyzer .
t
Place the calibration weight on the pan support .
y
Calibration starts automatically .
When calibration is complete, a report appears on the display . The report gives
the deviation detected during calibration . You are asked whether the weighing
system should be adjusted . After adjustment the report shows the result of the
adjustment .
t
Close the report: Select
<
.
y
The moisture analyzer is now calibrated . If you have also carried out an
adjustment after calibration, the moisture analyzer is also now adjusted .
CAL-Extern
START
Please set balance to zero
G - 0.401 g
CAL-Extern
Please put on required weight
Close hood
Stat - 50,000 g
Содержание MA37
Страница 65: ...65 Declaration of Conformity Operating Instructions Moisture Analyzer MA37 Declaration of Conformity ...
Страница 66: ...66 Certificate of Compliance Operating Instructions Moisture Analyzer MA37 Certificate of Compliance ...
Страница 67: ...67 Certificate of Compliance Operating Instructions Moisture Analyzer MA37 ...
Страница 68: ...68 Operating Instructions Moisture Analyzer MA37 Certificate of Compliance ...