Gentle drying:
For gentle drying, you need to enter the final temperature . The maximum set
temperature is 160°C . The target temperature is reached more slowly and the
measurement time is extended . Gentle drying is recommended for sensitive samples .
Standby Temperature
Temperature set to defined value when the sample chamber is closed .
Heating only occurs when the analyzer is in a measurement cycle . The standby
temperature is not the same as the measurement temperature .
The “Timeout Standby Temp .” can be set to 1 to 24 hours and is preset to 2 hours .
To change the setting, see Chapter “System Settings (Setup Menu)” .
Initial Weight
The target weight (as a guide for applying a sample to the analyzer – not as a
starting condition) can be entered by the user . The tolerance range can be freely
selected, e .g ., ± 0 .5 g .
Print Lines
When print lines are activated (1 or 2 lines with 20 characters each), additional
information about the method, entered by the user, appears on the report and
printout .
Filter
The user has the option of selecting between 0, 1, or 2 filters . During taring a
message appears indicating that the configured number of filters must be placed
with the sample pan for taring .
Target Value
The user can enter a target value with a tolerance range . During analysis a progress
bar appears, indicating the progress of the measurement . At the end of the analysis,
you can see whether the value is within the tolerance range .
Standardtrocknung
T [°C]
t [s]
18
Operating Instructions Moisture Analyzer MA37
Basic Principles
Содержание MA37
Страница 65: ...65 Declaration of Conformity Operating Instructions Moisture Analyzer MA37 Declaration of Conformity ...
Страница 66: ...66 Certificate of Compliance Operating Instructions Moisture Analyzer MA37 Certificate of Compliance ...
Страница 67: ...67 Certificate of Compliance Operating Instructions Moisture Analyzer MA37 ...
Страница 68: ...68 Operating Instructions Moisture Analyzer MA37 Certificate of Compliance ...