System Settings (Setup Menu)
Operating Instructions Moisture Analyzer MA37
Device Information
The manufacturer, model, serial number and software versions of the moisture
analyzer are displayed here .
Calibration/Adjustment
Calibration/Adjustment
MA performance test
Weighing system adjustment
Calibration report
The settings for calibration and adjustment of the moisture analyzer can be defined
here .
Under
MA performance test
:
Test functions for different temperature ranges are available here . You can only use
these functions with the ReproEasy pad (accessory) .
Under
Weighing system adjustment
:
t
Select the option of your choice .
–
CAL-Extern
: Carry out calibration with an external calibration weight
(see Chapter “Calibration and Adjustment”) .
–
Weighing only
: Use the weighing function to use the moisture analyzer
as a simple lab balance (see Chapter “Calibration and Adjustment”, section
“Weighing”) .
Under
Calibration report
:
Here you can access the most recent calibration reports for the MA performance
tests and weighing system adjustments . Up to 99 reports are saved per day .
Calibration reports that are at least 30 days old are deleted .
Printout
You can configure the settings for printing and data output here . Some of
the settings depend on the configuration of the USB interface (see Chapter “USB
Interface” on page 30) .
Under
Printout measurement
:
t
Select the settings for printing out the results on the connected lab printer .
–
Printout measurement
: The printout can be activated and deactivated
during the measurement here .
–
Interim results
: If the printout is activated, you can also activate the
output of intermediate values during the measurement here .
–
Interval [sec]
: Enter the desired interval in seconds for the output of the
intermediate results .
Under
Printout and report format
:
t
Configure all settings for the output format .
–
Result only
: Only the result of the moisture analysis is output .
–
With GLP and method parameters
: The GLP data and parameters used for
the moisture analysis are also output .
Device information
Manufacturer:
Sartorius
Model:
MA37
Serial number:
0034002458
Version BAC:
00-53-01.01 CN:1701
Version APC:
01-80-01.05 CN:0000
Version DOC:
07-40-03.00 CN:0000
Calibration report
MA performance test
Weighing system adjustment
Printout
Printout measurement
Printout and report format
Содержание MA37
Страница 65: ...65 Declaration of Conformity Operating Instructions Moisture Analyzer MA37 Declaration of Conformity ...
Страница 66: ...66 Certificate of Compliance Operating Instructions Moisture Analyzer MA37 Certificate of Compliance ...
Страница 67: ...67 Certificate of Compliance Operating Instructions Moisture Analyzer MA37 ...
Страница 68: ...68 Operating Instructions Moisture Analyzer MA37 Certificate of Compliance ...