![Sartorius MA37 Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/sartorius/ma37/ma37_operating-instructions-manual_1199627059.webp)
59
Care and Maintenance
Operating Instructions Moisture Analyzer MA37
4
) Clean the heating element and temperature sensor with a weak solvent such as
ethanol .
5
) If necessary, pull the air extractor grate out of the heating module . Then clean
the heating module and grate in a dishwasher .
6) Once cleaning is complete, reassemble the heating module and reinstall in the
analyzer, following the instructions in reverse order . The heating module must
snap into place on the analyzer .
Cleaning the Bottom of the Sample Chamber
Danger of Burns
:
Parts of the sample chamber may be extremely hot .
Wait until the bottom of the sample chamber has cooled completely .
Removing the Sample Chamber Bottom
1) Unlock the “Locking socket (bayonet)” as in general view with a suitable coin .