Measuring range:

min.

max.

1 m

3

/h

16 m

3

/h

Max. error of the measuring system:

± 2% (1 m

3

/h-16 m

3

/h)

Measuring signal:

2500 pulses =1 m

3

Measuring range of the frequency 
measurement:

0.6 Hz – 12 Hz

Accuracy:

± 0.005 Hz

Automatic shutoff after
1 minute at a frequency of:

< 0.4 Hz

RS232 serial interface:

19200 baud, 8 data bits, 1 stop bit, no parity

Current consumed by 
measurement electronics

:

Quiescent current: 

120 μA

Operating current: 

225 mA

Battery pack:

NiCd battery pack 9.6 V 

Capacity: 1000 

mAh

Dimensions of battery pack:

96 

+

28 

+

28 mm

Weight of the battery pack:

205 g

Hours of operation 
with fully charged battery pack:

approx. 4 h

Time it takes to load a completely 
discharged battery pack:

approx. 10 h

Battery monitoring*: 
Automatic charging module

When the calibration unit is supplied by
mains power, the battery pack is charged
through an automatic charging module.
Once the preset end-of-charging voltage
(10.6 V) is reached, the charging current 
(100 mA) is automatically reduced. This pre-
vents the battery pack from overcharging.

Power supply requirements:

14.5 V DC 530 mA

Allowable ambient temperature
of the calibration unit:

0°C .... +40°C

Type of protection:

IP 40 during operation with filter holder, 
connector and PVC hose

Numerical display of flow rate on 
the LCD display (in m

3

/h and l/min)

Alphanumerical display of 4 

+

20 pixels

Weight of the calibration unit 
without accessories:

Approx. 11 kg

Dimensions:

Length = 300 mm (without filter holder) 
Width = 390 mm with handles 
Height = 182 mm min., 

200 mm max. (adjustable feet)

Type of connector on the inlet 
and outlet ports for the filter holder
and MD8 airscan/AirPort MD8:

Bayonet-design quick connect

* The calibration unit can be run permanently on the battery charger without 

overloading the rechargeable battery pack. The unit can also be used when 
recharging (run on mains power supply). 
We recommend leaving the calibration unit connected to the battery charger |
power supply unit when it is not being used to prevent the rechargeable 
batteries from becoming completely discharged.

Important Notes

Caution! 

– Do not expose the calibration unit

to strong vibrations.

– To prevent contamination with 

airborne particles, do not operate
the calibration unit without a filter
in place.

– The maximum flow rate is 16 m

3

/h.

This rate may not be exceeded over
long time periods. 

– Only connect the calibration unit 

to MD8 airscan | AirPort MD8 air
samplers with a regulated air flow
rate. Never connect the calibration
unit to an air sampler with unknown
flow rate or you may possibly over-
load the flow rate measuring
device.

– The unit can be operated independ-

ent of mains power supply. When
operating the unit on mains power
supply however, make sure that
voltage rating is correct. If the unit
is operated independently of mains
power most of the time, it should
not be recharged using the power
supply unit until the prompt
appears in the display. This mini-
mizes a drain on battery capacity
caused by the battery’s memory
effect. When the unit is permanent-
ly operated on the mains power
supply, it is recommended to leave
the unit connected to the mains
power supply even when it is turned
off because the built-in automatic
charging system switches over 
to a very low charge conservation 
current during these times.

– Do not expose the unit to extreme

fluctuations in temperature or
humidity during operation or stor-
age. If fluctuations in temperature
and humidity cannot be avoided,
allow 2 hours for acclimatization.

– Use other than for the intended

purpose, or improper operation,
frees the manufacturer from all 
liability.

3

Specifications

!

Содержание 16756

Страница 1: ...Operating Instructions Bedienungsanleitung Calibration Unit 16756 Kalibriereinheit 16756 85030 523 26 Vers 11 2011...

Страница 2: ...alibration Unit 3 3 Specifications 4 4 Equipment Supplied 5 5 Operation of the Calibration Unit 5 6 Calibration of the MD8 airscan 7 7 Calibration of the AirPort MD8 8 8 Care and Maintenance 9 9 Infor...

Страница 3: ...ont panel Fig 1 see a and an air outlet port on the rear panel see b for attaching an 17085 hose for connection to the air sampler bayonet design quick connect Located on the right side Fig 1 c is the...

Страница 4: ...he battery charger without overloading the rechargeable battery pack The unit can also be used when recharging run on mains power supply We recommend leaving the calibration unit connected to the batt...

Страница 5: ...one on the MD8 airscan AirPort MD8 air sam plers See the operating instructions for the MD8 airscan AirPort MD8 air samplers regarding preparation of the filter holder placing the filter in the filte...

Страница 6: ...uring calibration no filter holder is attached to the air sampler Attach the hose to the end of the connector Fig 4 and attach the 17659 connector hose calibration unit to the free end of the hose Now...

Страница 7: ...alue set this value on the MD8 airscan by press ing or Then confirm and store it with Enter e If this is not the last calibration point the next one is automatically activated and processed as describ...

Страница 8: ...flow rate 7 3 Calibration Procedure After connecting the calibration unit and pressing the Start button on the calibration unit the message Sartorius MD8 Calibrating System start air sampler appears i...

Страница 9: ...in your regular household waste such equipment is manu factured from high grade materials and can be recycled and reused The European Directive 2002 96 EC on Waste Electrical and Electronic Equipment...

Страница 10: ...ax 44 1372 726171 In countries that are not members of the European Economic Area EEA or where no Sartorius affili ates subsidiaries dealers or distrib utors are located please contact your local auth...

Страница 11: ...not differentiate between still utilizable and or repairable used electrical and electronic equipment and no longer utilizable and or repairable used electrical and electronic equipment or between was...

Страница 12: ...er tebeschreibung 13 3 Technische Daten 14 4 Lieferumfang 15 5 Inbetriebnahme des Ger tes 15 6 Kalibrierung des MD8 airscan 17 7 Kalibrierung des AirPort MD8 18 8 Pflege und Wartung 19 9 Entsorgungshi...

Страница 13: ...er das Ger t unter Umgehung der Abschaltautomatik ausschaltet die Kalibriereinheit besitzt keinen Ausschalter die Kalibriereinheit schaltet sich nach 1 Minute selbst t tig ab Auf der Oberseite der Kal...

Страница 14: ...Max Fehler des Messsystems 2 1 m3 h 16 m3 Messsignal 2500 Impulse 1 m3 Messbereich der Frequenzmessung 0 6 Hz bis 12 Hz Genauigkeit 0 005 Hz Automatische Abschaltung nach 1 Minute bei Frequenz 0 4 Hz...

Страница 15: ...Filters und Verschlie en des Filterhalters erfolgen entsprechend der Bedienungsanleitungen f r die MD8 airscan AirPort MD8 Lufkeimsammelger te in denen auch die Handhabung Gebrauch der Gelatinefilter...

Страница 16: ...60 Sekunden sollte jetzt mit der Kalibrierung begonnen werden Geschieht dies nicht schal tet sich das Ger t nach 60 Sekunden ab Durch erneutes Dr cken der gr nen Starttaste kann das Ger t wieder einge...

Страница 17: ...end best tigen und spei chern Sie den Wert durch Bet tigen der Enter Taste e Wenn dies noch nicht der letzte Kalibrierwert ist m ssen Sie jetzt den n chsten Kalibrierwert wie unter a d beschrieben anf...

Страница 18: ...librierung Nach Anschluss der Kalibriereinheit und Dr cken der Starttaste an der Kalibriereinheit erscheint Sartorius MD8 Calibrating System start air sampler Am AirPort MD8 wird nun aus dem Men punkt...

Страница 19: ...m ll denn sie sind aus hochwertigen Materialien hergestellt die recyclet und wieder verwendet werden k nnen Die europ ische Richtlinie 2002 96 EG WEEE fordert die elektrischen und elektronischen Ger t...

Страница 20: ...34 91 3589623 Ungarn Sartorius Stedim Hung ria Kft Kagyl u 5 2092 Budakeszi Telefon 36 23 457 227 Fax 36 23 457 147 In den L ndern die keine Mitglieder des Europ ischen Wirtschaftsrau mes sind oder i...

Страница 21: ...u betr chtlichen Verz gerungen in der Abwicklung Bei der R cknahme wird nicht zwischen noch verwendungs bzw reparaturf higen Elektroaltger ten und nicht mehr verwendungs bzw reparaturf higen Elektroal...

Страница 22: ...ermany All rights reserved No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius Stedim Biotech GmbH The status of th...

Отзывы: