Press ZERO key twice while this message is
indicated on the screen (about 20 seconds).
Caution
This message is indicated on display for about
20 seconds. If no ZERO key is pressed when
the reading is on screen, 「Zeroing」 is
automatically interrupted and returns to the
Press ZERO key . -- ①
beginning of the process Try again
The buzzer beeps.
zeroing from the beginning if necessary.
Press
ZERO key. -- ②
The buzzer beeps.
Press the probe to 「Zero Plate」 when on the
reading for 20 seconds.
The buzzer beeps.
The reading shows that it is process of zeroing
adjustments
Measured numerical value
● Repeat a measuring process 2~10 times by pressing
the probe to 「Zero plate」.
Remove the probe from the 「Zero Plate」
(A measurement value is displayed whenever a probe
is pressed)
◆ When a measuring process reaches the 10
th
time, the
buzzer beeps and zeroing automatically finishes without
pressing ZERO key.
-19-
OFFSET.
To continue.
press ZERO twice.
【ZERO Cal.】
Measure uncoated
base metal
several times.
Afterwards,
press ZERO .
ZERO cal (Fe)
1
μm
Cal. No. 7