6)Calibration standard (CAL)
●
Prepare 「Zero Plate」 used in 「Zeroing」.
●
Prepare 「Thickness Standard」 which thicknesses is thicker or as thick as a measuring film.
●
Place 「Thickness Standard」 on 「Zero Plate」.
●
Press △ key or ▽ key.
The buzzer emits a beeping sound. It stops the LOCK function of LOCK/DELETE key
and changes to one data
“
deletion
”
function.
The buzzer beeps whenever the probe is pressed. The reading shows that it is in process of
Calibration standard.
The reading shows that the unit is set to measure
on the substrate material of Ferrous by connecting
a probe of the 「Fe」series.
Measured numerical value
In case the measured value abnormally deviates from the
thickness standard, a press of LOCK/DELETE key deletes
the latest entry and shows “--“ when all data were deleted.
● Repeat measuring operations 2~10 times by pressing probe
to 「Thickness standard」 on the「Zero plate」.
(The measured value is indicated whenever the probe is
pressed.)
Press or key after finsihing key operation.
◆ When the measuring process reaches the 10
th
time,
the buzzer beeps to take the next page screen.
◆
When the measuring process reaches less than 10
th
time,
being left doing nothing for 20 seconds, the buzzer beeps,
beeps to move to the next page screen.
-21-
【Cal.w.foil】
Measure STD-foil
on base metal
several times.
Afterwards,
Press or .
cal.w.fo
(Fe)
102
Cal. No. 7 ■
μm
△
▽
Caution
This message is indicated for approx. 20 sec.
ithout new entry during this period by
pressing △ or ▽ key, the unit
automatically returnato the begging from
where to start again for calibration.
△
▽