2nd Edition of TEMP2000_Series IM : July. 16. 2010 Page 48 / 87
▶
Following is the setup items.
①
Set Up, Down Signal Relay and Operating Deviation.
☞
Status Lamp and Relay run same way at Run Screen.
☞
For Program Run
-
Relay is ON until [Target Setting Point(TSP) – Set Temperature(Deviation)].
②
Soak Signal Relay and Soak Time.
☞
For Fix Operation → It displays status lamp on the screen but there is no relay output generated.
☞
For Program Operation
-
Status Lamp in the screen and Relay is ON until [Soak Seg Run Time(SEG TIME) – Set Temperature
(Soak Time)].
③
Set 1st, 2nd Freezer Run Signal Relay and Delay Time.
④
Set Error Signal Relay and Keep Time.
☞
Assigned Relay becomes “ON” when error occurs during Stop or Run.
☞
Relay maintains “ON” state for preset Keep Time, and becomes “OFF” after Keep Time.
☞
Relay maintains “ON” state when error is not recovered during Keep Time, and becomes “OFF” when
the error is recovered.
⑤
Set User Button Relay.
☞
Use of Button is set at [13. System Initial Setup].
☞
Corresponding Button is displayed on screen when the User Button is set.